Design and Performance Evaluation of Partial Geometry Configurations of Total-Body Pet: A Monte Carlo Study Publisher



Rezaei H1, 2 ; Sheikhzadeh P3 ; Ghafarian P4, 5 ; Ay MR1, 2
Authors

Source: Iranian Journal of Nuclear Medicine Published:2024


Abstract

Introduction: This study proposes various designs of partial-ring geometry total-body positron emission tomography (PET) scanners which maintain an extended axial field of view (AFOV) while remaining cost-effective, without sacrificing image quality. Methods: The total-body uEXPLORER PET scanner was simulated using the GATE Monte Carlo code. Partial-ring PET configurations were simulated, reducing the number of detectors by 25% in both the transaxial and axial directions. The sensitivity, noise equivalent count rate (NECR), and image quality of the partial-ring configurations were compared to those of the full-ring configuration. Results: The system sensitivity of partial-ring designs was 0.61 and 0.77 times that of the full-ring detector PET for detector reduction in the axial and transaxial directions, respectively. Notably, despite the higher sensitivity, the 10 mm sphere could not be detected in the transaxial reduction mode. When considering all spheres except the 10 mm one, the mean relative error of contrast recovery (CR) for the same time duration was 2% for detector reduction in the axial direction, whereas it increased to 9% for reduction in the transaxial direction. Conclusion: The results show that reducing the number of detectors in total-body PET scanners can significantly lower manufacturing costs associated with long AFOV PET systems. Although the reduced detector configuration resulted in lower sensitivity, the image quality of the NEMA phantom was found to be superior in the axial reduction mode compared to the transaxial reduction mode. Additionally, the image quality exhibited only a slight difference from that of the full-ring scanner. Copyright © 2024 The Authors. Published by Tehran University of Medical Sciences.
2. Accurate Monte Carlo Modeling and Performance Evaluation of a Total-Body Pet Scanner, 2021 IEEE Nuclear Science Symposium and Medical Imaging Conference Record# NSS/MIC 2021 and 28th International Symposium on Room-Temperature Semiconductor Detectors# RTSD 2022 (2021)
Experts (# of related papers)
Other Related Docs
6. A Generic, Scalable, and Cost-Effective Detector Front-End Block for Pet, 2017 IEEE Nuclear Science Symposium and Medical Imaging Conference# NSS/MIC 2017 - Conference Proceedings (2018)
8. A Multi-Purpose Clinical Pet Scanner With Dynamic Gantry Design, 2021 IEEE Nuclear Science Symposium and Medical Imaging Conference Record# NSS/MIC 2021 and 28th International Symposium on Room-Temperature Semiconductor Detectors# RTSD 2022 (2021)
15. Design, Optimization and Performance Evaluation of Bm-Pet: A Simulation Study, Nuclear Instruments and Methods in Physics Research# Section A: Accelerators# Spectrometers# Detectors and Associated Equipment (2019)
20. Validation and Evaluation of a Gate Model for Mammi Pet Scanner, Iranian Journal of Nuclear Medicine (2020)