Tehran University of Medical Sciences

Science Communicator Platform

Stay connected! Follow us on X network (Twitter):
Share By
Introducing a Dedicated Cardiac Spect Imaging System: Prospect Publisher



Fard BT1 ; Sabet KA2 ; Ay M1, 3
Authors

Source: Frontiers in Biomedical Technologies Published:2019


Abstract

Purpose: Cardiac SPECT imaging is widely used for many clinical practices such as the diagnostic of coronary artery diseases (CAD) and Myocardial Perfusion Imaging (MPI). In this work, we introduced the ProSPECT system, a dedicated cardiac SPECT system with open-gantry design. Materials and Methods: In this study we assessed the performance of the system based on NEMA-NU1-2007 standards. The ProSPECT system was characterized by measurement of planar, tomographic and clinical parameters. Results: Planar measurements showed 7.6 mm and 8.9% of spatial resolution and energy resolution at 140 keV, respectively. The tomographic resolution at 250 mm radius of rotation was 12.3 mm. Also, the maximum LOR and COR error for the tomographic imaging are obtained 1.1 mm and 1.8 mm, respectively. The clinical images that obtained using the system were confirmed by nuclear medicine physician expert. Conclusion: It concluded that the ProSPECT imaging is qualified for cardiac SPECT imaging. © Tehran University of Medical Sciences
Other Related Docs
17. Deep Learning-Based Low-Dose Cardiac Gated Spect: Implementation in Projection Space Vs. Image Space, 2021 IEEE Nuclear Science Symposium and Medical Imaging Conference Record# NSS/MIC 2021 and 28th International Symposium on Room-Temperature Semiconductor Detectors# RTSD 2022 (2021)