Tehran University of Medical Sciences

Science Communicator Platform

Stay connected! Follow us on X network (Twitter):
Share By
Static Coil Design Considerations for the Magnetic Resonance Imaging Publisher



Shiravi M1 ; Ganji B1 ; Shiravi A2
Authors

Source: International Journal of Engineering Transactions C: Aspects Published:2019


Abstract

One of the main challenges in developing magnetic resonance imaging (MRI) systems is to create a static coil that needs to generate magnetic field density along with the characteristics of optimal homogeneity and magnitude size. To do this, two N42 Block PMs are used and the iron core is designed and optimized in accordance with the dimensions of PM pieces using ANSYS Maxwell software. Then, all iron parts are lathed, the yoke pieces and pole spacers are welded. In addition, PM and pole pieces are installed. Finally, measurement is done by Lutron to evaluate the static coil performance. © 2019, Materials and Energy Research Center. All rights reserved.
Related Docs
1. A Novel Wearable Rf Head Coil for High Resolution 7T Magnetic Resonance Imaging, 14th European Conference on Antennas and Propagation# EuCAP 2020 (2020)
Experts (# of related papers)